Cleaning of the reference junction

by Andy
(US)

It's not only the glass membrane of the pH electrode that require frequent cleaning. Reference junction fouling or plugging remains the primary challenge to pH measurement
in dirty applications.


Clogged reference junction treatment:

It's recommended to remove the electrolyte and replaced with fresh electrolyte. The electrode should then be heated to 60-80°C in 3M KCl and allowed to cool in the same solution for approximately 10 minutes. Rinse and put into a pH 4 buffer for at least 1/2 hour.

Comments for Cleaning of the reference junction

Average Rating starstarstarstarstar

Click here to add your own comments

May 31, 2010
Rating
starstarstarstarstar
Reference junction cleaning
by: Will

In many cases, the junction can be cleaned with aggressive alkaline cleaners (for oil plugging) or dilute acids (for salt deposits) or a combination of both.

Click here to add your own comments

Join in and write your own page! It's easy to do. How? Simply click here to return to Electrode cleaning tips and advice.



Alibris: Books, Music, & Movies

SuppliesShops.com - Click Here for ALL Your Office Supply Needs!


Favorite pages:


acids and bases
What is an acid? What is a base? If you want to know, read here!

calculating ph
Do you know what pH stand for? If not, click here!

electrode design
The pH electrode consists of the following parts:
. . read here!
offset error
Real-world pH electrodes are not ideal,
read why.

definition of ph
What is the definition of pH? See the answer. . .

nonglass ph electrode
Is it not possible to use a glass electrode for your application? Use a non-glass pH electrode.

ph measurement
Learn how the temperature affects your pH measurement.


Download the new release ISO 14001:2015