pH Electrode cleaning solution

Optimal electrode cleaning procedures are dependent upon the type and the thickness of the coating. The glass membrane should only in extreme cases be mechanically cleaned as abrasion can lead to permanent damage. Even wiping or drying the glass membrane can lead to faults in the measurement function. Hard coatings should be chemically removed. A good advice is to clean the pH sensor regularly.

The first simple action

In most cases, rinsing the electrode after measurement with clean water is just enough.

Electrode cleaning solution

If, after rinsing in water, your pH electrode still not works properly, you need to use a cleaning agent. The cleaning agent used basically depends on the type of contamination. For cleaning agents, check with your pH electrode supplier. Your pH electrode can be cleaned with whatever chemical that can dissolve the coating, provided the chemical will not attack the electrode's materials of construction. However, a good rule is to minimize the exposure of the pH sensor to the cleaning solution and only immerse the glass membrane or the glass membrane and the diaphragm.

Tips and advice on which cleaning agent to use:

• General

Soak in 0.1M HCl for half an hour.


• Oil and grease

Warm water with some detergent solution is probably sufficient. An alternative is to rinse with an ethanol solution.


• Protein

Soak in 1% pepsin and 0.1M HCl solution for about 1 hour.


• Sulfides

For most pH electrodes a measuring solution containing sulfides will result in silver sulfide contamination of the diaphragm that is difficult to dissolve. Soak in 0.1M Thiourea/HCl solution for 15 min to 1 hour or until the diaphragm is totally bleached.


• Inorganic

Soak in 0.1M tetrasodium EDTA solution for 15 minutes. An alternative is to use commercial glass cleaning solution (not strongly acidic).


• Acid or alkaline deposits

Alkaline deposits can be removed with weak acid solutions (0.1 molar HCl) or vinegar. Acidic deposits can be removed with mild alkaline solutions (0.1 molar NaOH.)

Soak the glass membrane for a few minutes. Keep the acid or alkaline solvent away from the liquid junction. If not, hydrogen ions (acid solvent) or hydroxide ions (alkaline solvent) will diffuse into the junction and causing the liquid junction potential and the pH reading to drift. It may take hours or days for the reading to stabilize.

After treatment

After any of these electrode cleaning procedures, rinse thoroughly with water and soak the electrode in storage solution overnight. If the electrode is refillable, drain and refill the reference solution. Always recalibrate the pH electrode after cleaning. If the electrode response is slow or the electrode does not calibrate correctly, you maybe need to restore the hydrated glass layer.

Share your point of view with others!

Take the opportunity to share your knowledge and experiences with others.

Please, use the form below to publish your contribution about the subject "pH Electrode cleaning". Before your page is published, it will be reviewed to make sure it applies to the submission guidelines for this website.

Don't forget to read what others have written, maybe you have a comment to add.



What Other Visitors Have Said

Click below to see contributions from other visitors to this page...

Cleaning of the reference junction 
It's not only the glass membrane of the pH electrode that require frequent cleaning. Reference junction fouling or plugging remains the primary challenge …

Click here to write your own.




SuppliesShops.com - Click Here for ALL Your Office Supply Needs!


Favorite pages:


acids and bases
What is an acid? What is a base? If you want to know, read here!

calculating ph
Do you know what pH stand for? If not, click here!

electrode design
The pH electrode consists of the following parts:
. . read here!
offset error
Real-world pH electrodes are not ideal,
read why.

definition of ph
What is the definition of pH? See the answer. . .

nonglass ph electrode
Is it not possible to use a glass electrode for your application? Use a non-glass pH electrode.

ph measurement
Learn how the temperature affects your pH measurement.




Download the new release ISO 14001:2015